ICA 系列高压电源 (Featured)

ICA 系列

  • 多路输出电子枪电源
  • 快上升沿脉冲阴极电流
  • 通过以太网远程设定
  • 可针对 OEM 应用进行定制

*注意:所有规格如有变更,恕不另行通知。有关最新修订,请参阅本数据表的英文 PDF 版本

CE、ROHS、UL 标识

喷油器控制组件 (ICA) 电源

斯派曼注射器控制集合(ICA)多路输出高压电源是使用来供应适当的电极电压到一个三极管电子枪,为了注射电子到一个直线加速器(LINAC)。由此产生的高能电子脉冲用于为应用来创建高能的X 射线,比如货物检查和癌症治疗。

此多路输出机架式电源通过电子枪提供阴极加热器和栅极电压。此阴极电流可以被脉冲到500Hz。此加热器和栅极电源浮在阴极电压电势上,通常一直到-15kV。注射器电流是通过编程栅极脉冲电压控制的。远程控制和监测是通过标准的以太网协议。在前面板上的LED 指示器提供基本的输出和故障状态,包括阴极电缆和电弧故障。

此ICA 系列可以为特殊的系统需求定制。

典型应用

  • 货物检查、癌症治疗

规格

(Ref. 128129-001 REV. C)

Input Voltage:

100-240Vac, 50-60Hz

Cathode:

Output Voltage:

0 to -15kV

Pulsed Cathode Current:

200mA to 1500mA

Mean Cathode Current:

5.5mA (max)

Current Pulse Width:

Up to 6μs (90% to 90%)

Current Pulse Frequency:

Up to 500Hz

Current Risetime:

200ns (typical)

Heater:

Output Voltage:

0 to -6.5Vdc (wrt Cathode)

Current:

3A

Resistance:

0.7ohm

Grid (fixed):

Voltage:

-60 to -135Vdc (wrt Cathode)

Current:

250mA

Pulse Voltage:

0 to 340Vdc (wrt Cathode)
Injector Current is regulated by programming
the Grid Pulse Voltage

Connections:

Front Panel:
Ethernet (RJ45 jack)
Interlock (24V nominal, 9 pin D-sub male pin plug)
Pulse Input (BNC)

Rear Panel:

AC Power In (IEC320 C14 male pin receptacle)
HV Out (3 pin terminal block)

Interlock:

24V must be present to enable output high voltage cathode. It
is intended to connect to the system to disable the ICA in case
any system interlocks have not been satisfied.

Operating Temperature:

10°C to +40°C

Storage Temperature:

-40°C to +70°C

Humidity:

30% to 60% relative humidity, non-condensing

Weight:

15lbs. (6.8kg)

Regulatory Approvals:

The ICA shall be designed in accordance with IEC 60601-1
(Edition 3) Requirements. Certification shall be by agreement.
The ICA shall be UL 94V-0 and RoHS compliant.

INDICATOR COLOR CONDITION Illuminated When...
Trigger Present Green Input trigger is present.
AC Power Green AC supply voltage is present.
Interlock In Green Interlock in signal is present.
Arc Fault Red Arc fault signal is active.
Cathode Current Fault Red Cathode current fault signal is active.
Cathode Voltage On Amber Cathode high voltage is present.

 

表和图

ICA Injector Control Assembly for Cargo Screening & Cancer Therapy - Front View

ICA Injector Control Assembly for Cargo Screening & Cancer Therapy - Top View

ICA Injector Control Assembly for Cargo Screening & Cancer Therapy - Back View