XRB80HR Monoblock® 80kV, 100W X射线发生器 (featured)

XRB80PN100HR Monoblock® X射线发生器

  • 集高压电源、灯丝电源、X射线管,射线束端口和电子控制器件于一身
  • 体积小,重量轻
  • 通用输入,通过内部 EMI 滤波器校正功率因数
  • 可以安装在任何物理方位
  • 模拟监测接口和标准的 RS-232 数字编程和监测接口
  • 数据录入和软件控制X光管老化(仅灵活的控制选项)

*注意:所有规格如有变更,恕不另行通知。有关最新修订,请参阅本数据表的英文 PDF 版本

3 Year WarrantyCE RU Marking

80kV,100W 集成 X 射线源

  斯派曼 XRB80PN100HR(高可靠性)Monoblock® X 射线源是专为 OEM 应用而设计。输出电压为 80kV,功率为 100W,为内部 X 射线管供电。产品具有通用输入、小的外观尺寸、标准的模拟接口和 RS-232 数字接口等特点,使得此 Monoblock® 能更简单集成到您的 X 射线系统中。此 XRB80PN100HR 提供扇形(标准的)或锥形(可选择的)射线束几何形状。专有的发射控制电路提供了卓越的 X 射线管电流管理,同时具有出色的稳定性能。XRB80PN100HR 专为长的现场使用寿命而设计,并提供 3 年保修期。

典型应用:

  • X 射线扫描:厚度测量、食品检验、液位确认、包裹检查

规格

(Ref. 128118-001 REV. N)

X-Ray Characteristics:

Focal Spot:

0.8mm (IEC 336) standard

0.5mm (IEC 336) optional

Beam Filter:

Ultem: 3.00mm ±0.15mm

Oil: 7.5mm ±0.25mm

Glass: 1.7mm ±0.2mm

Be: 0.8mm

Beam Geometry:

Fan: Standard. The beam angular coverage will be 80° with the beam plane perpendicular to the X-Ray tube axis and 20° wide (with a 2° tolerance)

Cone: Optional. 20° cone beam (with a 2° tolerance)

Input Voltage:

Power factor corrected input 0.98, 100-240Vac ±10% 50/60Hz, 2A maximum

X-Ray Tube Voltage:

Nominal X-Ray tube voltage is adjustable 40kV (±20kV) to 80kV (±40kV)

X-Ray Tube Current:

150uA to 2.00mA over specified tube voltage range (100W max.)

X-Ray Tube Power:

100W, maximum continuous

Voltage Regulation:

Line: ±0.05% of maximum output voltage over a ±10% change of nominal input line voltage

Load: ±0.1% of maximum rated voltage for 150uA to 2.00mA load change

Voltage Accuracy:

Voltage measured across the X-Ray tube is within ±2% of the programmed value

Voltage Risetime:

Standard: Ramp time shall be <500ms from 10% to 90% of maximum rated output voltage

Voltage Ripple:

0.5% peak to peak of maximum voltage for frequencies ≤1kHz

Emission Current Parameters

Current Regulation:

Line: ±0.05% of rated output current over a ±10% change of nominal input line voltage

Load: ±0.1% of rated output current for a change from 50% to 100% of rated output voltage

Current Accuracy:

Current measured through the X-Ray tube is within ±2% of the programmed value

Current Risetime:

Standard: Ramp time shall be <500ms from 10% to 90% of maximum rated current

Arc Intervention:

4 arcs in 10 seconds with a 100ms quench/100ms re-ramp = Shutdown

Filament Configuration:

Internal AC filament drive with closed loop filament emission control

Analog Monitoring Interface:

Ground referenced 0 to 9Vdc for all monitoring signals.

Relay contacts and open collector signals for other signals. See analog interface connector pin out table.

Digital Programming and Monitoring Interface:

The RS-232 interface allows for programming of kV, mA output and X-Ray enable. Provides monitoring for kV, mA output and oil temperature. Tolerance 3%. (with an additional 5μA offset at ≤10% mA programming)

Control Software:

A demo GUI is available for engineering evaluations

Operating Temperature:

0°C to +40°C

Storage Temperature:

-40°C to +70°C

Humidity:

10% to 95% relative humidity, non-condensing

Cooling:

X-Ray Tank: Customer provided 250 cfm external cooling fan as required to maintain oil temperature below 55°C.

Controller: Forced air via internal fan.

Input Line Connector:

3-pin Phoenix Contact 1829167. Mating connector provided with unit

Analog Interface Connector:

15 pin Male D connector provided with unit

Digital Interface Connector:

9 pin female D connector, female

Grounding Point:

M4 ground stud provided on chassis

Dimensions:

X-Ray Tank: 11.3˝L X 9.625˝W X 4.93˝H (287.02mm x 244.4mm x 125.2mm

Controller: 8.5˝L X 6.70˝W X 2.21˝H (215.9mm x 170.2mm x 56.1mm)

Weight:

X-Ray Tank: 36lbs (16.32kg)

Controller: 3.7lbs (1.68kg)

Orientation:

Can be mounted in any orientation.

X-Ray Leakage:

Not to be greater than 0.5mR/hr at 5cm outside the external surface.

Regulatory Approvals:

Compliant to EEC EMC Directive. Compliant to EEC Low Voltage Directive. UL/CUL recognized file E235530

Options:

CB Cone Beam
.5mm .5mm focal spot X-Ray tube
NF 80° x 10° Narrow Fan beam
RA Right Angle cable
SC Smart Controller

AC LINE POWER CONNECTOR— J1 THREE POSITION PHOENIX CONTACT

Pin Signal
1 Earth Ground
2 Line
3 Neutral

Mating connector provided with unit

RS-232 DIGITAL INTERFACE— J3 9 PIN FEMALE D CONNECTOR

Pin Signal Parameters
1 N/C No Connection
2 TD Transmit Data
3 RD Receive Data
4 N/C No Connection
5 SGND Signal Ground
6 NC No Connection
7 NC No Connection
8 NC No Connection
9 NC No Connection

XRB80HR ANALOG INTERFACE— J2 15 PIN MALE D CONNECTOR

Pin Signal Parameters
1 Power Supply Fault Output Open collector, 35V @ 10mA max. high = no fault
2 N/C No Connection
3 N/C No Connection
4 X-Ray On Lamp Relay Output Common, dry contacts, 30Vdc @ 1A, max
5 X-Ray On Lamp Relay Output Normally open, X-Ray ON = closed
6 mA Monitor Output 0 to 9Vdc = 0 to 100% rated output, Zout =10kΩ
7 X-Ray On Lamp Relay Output Normally closed, X-Ray ON = open
8 kV Monitor Output 0 to 9.00Vdc = 0 to 100% rated output, Zout =10kΩ
9 Signal Ground Ground
10 Signal Ground Ground
11 HV Interlock Return Input Connect to Pin 12 to close HV interlock
12 HV Interlock Output +15Vdc @ open, 5mA when connected to pin 11
13 X-Ray Enable Output +15Vdc @ open, 5mA when connected to pin 15
14 X-Ray Status Output Open collector, 35V @ 10mA max high = X-Ray OFF
15 X-Ray Enable Return Input Connect to pin 13 to enable X-Ray generation (for local enable)

LED INDICATORS

Indicator Signal Name Condition Illuminated When...
LED 1 OV High kV occurs
LED 2 UV Low kV occurs
LED 3 UC Low mA occurs
LED 4 OC High mA occurs
LED 5 ARC FLT Arc fault occurs
LED 6 OT Over temperature occurs
LED 7 X-RAY ON X-Rays are enabled
LED 8 PWR Power is ON

 

 

How to Order:
Standard: Part No.: XRB80PN100HR

Cone Beam Option

Part NO.: XRB80PN100HR/CB

0.5mm Focal Spot Option

Part NO.: XRB80PN100HR/.5mm

80° X 10° Narrow Fan Beam Option

Part NO.: XRB80PN100HR/NF

Cable Option:

Part NO.: XRB*)PN100HR/RA

 Smart Controller Option

Part NO.: XRB80PN100HR/SC

 

SMART XRB (only available with Smart Controller option)

The XRB80PN100HR with the Smart Controller (SC option) has two new digital features available: data logging and firmware controlled seasoning.

Data Logging:
Think of this as an “airplane black box”. The data logging captures data on fault events and non-fault events. Fault events will turn off the high voltage:

FAULT EVENTS
Temperature
High Current
High Voltage
Power Vault
Arc
Low Voltage
Watchdog
Interlock

The XRB80PN100HR stores data 620ms before the event, the event itself and for 620ms after the event. Data is recorded every 20ms (62 samples total) showing:

Anode kV
Total kV
Filament
Cathode kV
Total mA
Temperature

We also log non-fault events, these are changes in set points or state of the unit.

NON FAULT EVENTS
HV On
kV Set Point
Low Current
HV Off
mA Set point
Pre Heat Set Point

Fault event data is actual graphical data. Non fault event data is just stored as event type, data and timestamp. We also have a preventative maintenance fault, which throws a non-shutdown alarm if the X-Ray tube has been factory installed over 4 years ago or if over 15,000 hours of HV ON is logged.

Firmware Controlled Seasoning:
Every unit comes with an initial seasoning table, or customers can set their own. The XRB80PN100HR knows when the unit has been on, when it has been off, hours on the X-Ray tube, etc. As a preventative maintenance feature upon turn on, we review the data and suggest that a particular seasoning protocol be run based upon the actual usage history of the unit. Proper seasoning compliance of the X-Ray tube will help get the longest lifetime.

 

表和图

OPTIONS

CB Cone BEam

.5mm .5mm focal spot X-Ray tube

NF80° x 10° Narrow Fan beam

RA Right Angle cable

DIMENSIONS: in.[mm]

CONTROL UNIT

TOP VIEW

XRB80HR Monoblock® X-Ray Source (Image 1) 

 

 

SIDE VIEW

XRB80HR Monoblock® X-Ray Source (Image 2)

BOTTOM VIEW

XRB80HR Monoblock® X-Ray Source (Image 3)

DIMENSIONS: in.[mm]

SMART CONTROL UNIT

TOP VIEW

XRB80PN100HR Smart Control Unit Top View

SIDE VIEW

XRB80PN100HR Smart Control Unit Side View

BOTTOM VIEW

XRB80PN100HR Smart Control Unit Bottom View

 

DIMENSIONS: in.[mm]

GENERATOR TANK

TOP VIEW

XRB80HR Tank View

FRONT VIEW

XRB80HR Front View

SIDE VIEW

XRB80HR Side View

 

 

Frequently Asked Questions

Why Is Oil Insulation Used?
Do I Need to Ensure My Monoblock® Stays Cool? Why?
How Often Do I Need to Season My Monoblock® X-Ray Source? Why?
Application Notes AN-12 – The Benefit of Using a Current Source to Power X-Ray Tube Filament Circuits