XRBe80PN300 Monoblock® 集成 X 射线源

  • 紧凑、轻便
  • 通用输入,功率因数校正
  • 可以以任何物理方向安装
  • 模拟监控接口、标准RS-232数字接口和以太网
  • 数据记录和固件控制的 X 射线管干燥

*注:所有规格如有变更,恕不另行通知。 请参阅此数据表的英文 PDF 版本以获取最新版本。

80KV/300W Be Window Monoblock®

斯派曼的 XRBe80PN300 铍窗 Monoblock® X 射线源专为 OEM 应用而设计,以 300W 的功率水平为内部双极 X 射线管供电,电压为 80kV。通用输入、紧凑封装尺寸和标准 RS-232 数字接口等功能简化了 XRBe80PN300 与 X 射线系统的集成。它具有锥形光束几何形状。 专有的发射控制电路可对 X 射线管电流进行出色的调节,并具有出色的稳定性能。

典型应用

食品检测系统、填充水平确认系统、安全扫描系统、工业无损检测系统、厚度/电镀测量系统

规格

(Ref. 128132-001 REV. C)

Options: 

ST-TE, RA-SE, RA-TE Cable Options (see page 3 of the data sheet)

SPECIFICATIONS

X-Ray Characteristics:

Focal Spot: 0.8mm (IEC 336) standard 

Beam Filter: Be 0.8mm

Beam Geometry:

Cone: Standard. The beam angular coverage will be 40° with the beam plane perpendicular to the X-Ray tube axis

Input Voltage:

100- 240Vac, ±10%, 50/60 Hertz, .98 power factor

Input Current:

300W @ 6A 

X-Ray Tube Voltage:

Maximum 80kV

X-Ray Tube Current:

Maximum 5mA

X-Ray Tube Power:

300W

Voltage Regulation:

Line: Line:±0.05% of maximum output voltage over a ±10% change of nominal input line voltage

Load: ±0.1% of maximum rated voltage for 150uA to full rated load change

Voltage Accuracy:

Voltage measured across the X-Ray tube is within ±2% of the programmed value

Voltage Risetime:

Standard ramp time shall be <500ms from 10% to 90% of maximum rated output voltage

Voltage Ripple:

0.5% peak to peak of maximum voltage forfrequencies ≤1kHz

X-Ray Tube Voltage:

Maximum 80kV

Emission Current Parameters

Current Regulation:

Line: ±0.05% of rated output current over a ±10% change of nominal input line voltage

Load: ±0.1% of rated output current for a change from 50% to 100% of rated output voltage

Current Risetime:

Standard ramp time shall be <500ms from 10% to 90% of maximum rated current

Arc Intervention:

4 arcs in 10 seconds with a 100ms quench/100ms re-ramp = Shutdown

Filament Configuration:

Internal floating AC filament drive with closed loop filament emission control circuitry

Analog Interface:

Ground referenced 0 to 9Vdc for all monitoring signals. Relay contacts and open collector signals for other signals. See analog interface connector pinout table.

Digital Interface:

The RS-232 interface allows for programming of kV, mA output and X-Ray enable. Provides monitoring for kV, mA output and oil temperature.

Operating Temperature:

0°C to +40°C

Storage Temperature:

-40°C to +70°C

Humidity:

10% to 95% relative humidity, non-condensing

Cooling:

X-Ray Tank: Externally powered forced air cooling with oil pump and heat exchanger, 24Vdc @ 5A

Controller: Forced air via internal fan.

Grounding Point:

M5 ground female thread on tank
M5 ground stud on control chassis provided 

Dimensions:

X-Ray Tank: See drawings

Controller: See drawings

Weight:

X-Ray Tank:

81.5 lbs. (37kg)

Controller:

7 lbs. (3.18kg)

Orientation:

Can be mounted in any orientation.

X-Ray Leakage:

Not to be greater than 0.5mR/hr at 5cm outside the external surface. 

Regulatory Approvals:

Compliant to EEC EMC Directive. Compliant to EEC Low Voltage Directive. UL/CUL recognized file E235530.

AC INPUT—3 PIN PHOENIX CONTACT 1858772

PIN SIGNAL PARAMETERS
1 AC Input (high) 100-240Vac (high)
2 Ground Ground
3 AC Input (neutral) 100-240Vac (neutral)

 

DC INPUT FOR HEAT DISSIPATION UNIT—4 PIN AMP(210/350/500W) 206060-1

PIN SIGNAL DESCRIPTION PARAMETERS 
1 24Vdc Fan/Pump Power 24Vdc @ 5 amps 
2 24Vdc Return Fan/Pump Power

24Vdc @ 5 amp

3 N/C No Connection N/C
N/C  No Connection N/C 

It is critical to ensure that 24Vdc is supplied to pins 1 and 2

LED INDICATORS

INDICATOR SIGNAL CONDITION Illuminated When... LED COLOR
1 OV Error 6 High kV occurs Red
2 UV Error 5 Low kV occurs Red
3 UC Error 4 Low mA occurs Red
4 OC Error 3 High mA occurs Red
5 ARC Error 2 Arc fault occurs Red
6 OT Error 1 Over temperature occurs Red
7 X-Ray On X-Rays are being generated Green
8 Power On AC input power is present Green

 

ANALOG INTERFACE—10 PIN PHOENIX CONTACT 1792605

PIN SIGNAL PARAMETERS
1 X-Ray Interlock Enable Apply +24Vdc to enable interlock. Open/removal of +24Vdc will cause X-Ray generation to stop.
2 X-Ray Interlock Enable Return Ground reference for X-Ray Interlock.
3 Pin Removed N/C
4 kV Monitor 0-10Vdc = 0-100% rated output voltage. Zout = 10kΩ
5 Signal Ground Signal Ground
6 mA Monitor 0-10Vdc = 0-100% rated output current. Zout =10kΩ
7 Fault Signal Open collector, High (Open) = No Fault, 35Vdc @ 10mA maximum
8 HV ON Lamp Relay N/O N/O dry contacts. 50 volts maximum. Rated for 1 amp or less, 50mA nominal load
HV ON Lamp Relay C  Common dry contacts. 50 volts maximum. Rated for 1 amp or less, 50mA nominal load 
10  HV ON Lamp Relay N/C  N/C dry contacts. 50 volts maximum. Rated for 1 amp or less, 50mA nominal load 

RS-232 DIGITAL INTERFACE—9 PIN MALE D CONNECTOR 

PIN SIGNAL PARAMETERS
1 N/C N/C
2 Transmitted Data Conforms to EIA RS-232-C
3 Received Data Conforms to EIA RS-232-C
4 N/C N/C
5 Signal Ground Signal Ground
6 N/C N/C
7 N/C N/C
8 N/C N/C
N/C  N/C 

ETHERNET DIGITAL INTERFACE—RJ45 8 PIN FEMALE CONNECTOR 

PIN SIGNAL PARAMETERS
1 TX+ Transmit Data +
2 TX- Transmit Data-
3 RX+ Receive Data +
4 N/C No Connection
5 N/C No Connection
6 RX- Receive Data -
7 N/C No Connection
8 N/C No Connection

XRBD Cable Configurations

 

SMART XRB

The XRBD has two new digital features: data logging and firmware controlled seasoning.

Data Logging:

Think of this as an “airplane black box”. The data logging captures data on fault events and non-fault events. Fault events will turn off the high voltage:

FAULT EVENTS

Temperature
Arc
High Current
Low Voltage
High Voltage
Watchdog
Power Fault
Interlock

The XRBe80PN300 stores data 620ms before the event, the event itself and for 620ms after the event. Data is recorded every 20ms (62 samples total) showing:

Anode kV
Cathode kV
Total kV
Total mA
Filament
Temperature

We also log non-fault events, these are changes in set points or state of the unit.

NON FAULT EVENTS

HV On
HV Off
kV Set point
mA Set point
Low Current
Filament Limit Set point
Pre Heat Set point
Line Dip

Fault event data is actual graphical data. Non fault event data is just stored as event type, data and timestamp. We also have a preventative maintenance fault, which throws a non-shutdown fault if the X-Ray tube has been factory installed over 4 years ago or if over 15,000 hours of HV ON is logged.

Firmware Controlled Seasoning:

Every unit comes with an initial seasoning table, or customers can set their own. The XRBe80PN300 knows when the unit has been on, when it has been off, hours on the X-Ray tube, etc. As a preventative maintenance feature upon turn on, we review the data and suggest that a particular seasoning protocol be run based upon the actual usage history of the unit. Proper seasoning compliance of the X-Ray tube will help get the longest lifetime.

表和图

DIMENSIONS: in.[mm]

CONTROL UNIT

FRONTVIEW

 

XRBe80PN300 Control Unit Front View

   

BOTTOM VIEW

 

XRBe80PN300 Control Unit Bottom View

REAR VIEW

XRBe80PN300 Control Unit Rear View

SIDE VIEW

XRBe80PN300 Control Unit Side View

DIMENSIONS: in.[mm] 

TOP VIEW

 XRBe80PN300 Control Unit Top View

 

SIDE VIEW

 XRBe80PN300 Control Unit Side View 2

FRONT VIEW

XRBe80PN300 Control Unit Front View 2