100W X 射线发生器
uXHP 系列采用斯派曼特殊的高压封装和表面贴装制造技术,结合斯派曼专有的封装技术,生产出超紧凑尺寸的X 射线发生器模块。uXHP 给很多著名制造商的阴极接地X 射线管供电,特征:高压输出电压从0 至50kV/65kV/80kV,电流从0 至5mA,功率限制为100W。uXHP 使用闭环灯丝控制电路,提供高精度调节射线束电流。这个低噪声直流灯丝电源工作在0.3A 和3.5A 之间。提供严谨性调控、高稳定性和低纹波。uXHP 提供用户本地和远程模拟控制,去设置管电压、管电流和灯丝电流限制。提供标准的USB、RS-232 和以太网接口。
典型应用
给Varex、Kevex、Oxford、RTW、Superior 和Trufocus 的阴极接地X 射线管供电。
规格
(Ref. 128110-001 REV. K)
Specifications
Input:
+24Vdc ±1V; 7.75A maximum
Efficiency:
75%, typical
Output:
See model selection table on page 2
Voltage Control:
Local: Internal multi-turn potentiometer to set voltage
from 0 to full output voltage.
Remote: 0 to 10Vdc = 0 to 100% rated output voltage.
Accuracy: ±1%. ZIN = 10Mohm.
Emission Control:
Local: Internal multi-turn potentiometer to set voltage
from 0 to full output voltage.
Remote: 0 to 10Vdc = 0 to 100% rated output voltage.
Accuracy: ±1%. ZIN = 10Mohm.
Emission Control:
Local: Internal potentiometer to set beam current
from 0 to full output current.
Remote: 0 to 10Vdc = 0 to 100% rated output current.
Accuracy: ±1%. ZIN = 10Mohm. Filament limit and
filament preheat control capability is also provided.
Voltage and Current Monitors:
0 to 10Vdc = 0 to 100% rated output
Accuracy: ±1%. ZOUT = 1kohm
Redundant Voltage Monitor:
A redundant high voltage feedback divider where
0 to 10Vdc = 0 to 100% rated output is monitored via firmware.
The analog monitor signal can be provided by special order.
Temperature Coefficient:
0.01% per °C, voltage and current.
Stability:
0.05% per 8 hours after 1/2 hour warm-up.
Digital Interface:
RS-232, Ethernet and USB is standard
DC Filament Supply:
Ground isolated filament power supply allows actual tube
current feedback signal for monitoring accurate low X-Ray
tube current performance.
Current: 0-3.5A, adjustable limit
Voltage: 5.0V, maximum compliance
Environmental:
Operational: 0° C to +50° C
Storage: -40° C to +85° C
Humidity: 0% to 90%, non-condensing
Dimensions:
50/65kV Unit: 7.00”H x 3.07”W x 9.00”D
(177.80mm x 78.00mm x 228.60mm)
80kV Unit: 7.00”H x 3.07”W x 10.50”D
(177.80mm x 78.00mm x 266.70mm)
Weight:
50/65kV Unit: 8.5 lbs. (3.85kg)
80kV Unit: 10.0 lbs. (4.53kg)
Cooling:
User provided forced air cooling is required
Regulatory Approvals:
Compliant to EEC EMC Directive. Compliant to EEC Low Voltage Directive. UL/CUL recognized, File E227588. RoHS Compliant.
uXHP MODEL SELECTION TABLE
OUTPUT VOLTAGE | OUTPUT CURRENT | OUTPUT POWER | MODEL NUMBER |
---|---|---|---|
0-50kV | 0-5mA | 100W | uXHP50P100 |
0-65kV | 0-5mA | 100W | uXHP65P100 |
0-80kV | 0-5mA | 100W | uXHP80P100 |
CONTROL POWER INPUT CONNECTOR
PIN | SIGNAL | PARAMETER |
---|---|---|
1 | +24V Control Power | +24V @ 1A max |
2 | +24V Control Power Return | Power Ground |
FILAMENT CONNECTOR
PIN | SIGNAL | PARAMETER |
---|---|---|
1 | Filament Output | 0.3A to 3.5A @ 5V, max |
2 | Filament Return | Filament Return |
Note: The filament return wire cannot be grounded as this would short circuit
the tube return current monitoring to the uXHP. If grounding of the filament is
required, please consult the factory.
ANALOG INTERFACE CONNECTOR MALE 15 PIN MINI “D”
PIN | SIGNAL | PARAMETER |
---|---|---|
1 | Monitor Return | Signal Ground |
2 | Voltage Monitor | 0-10V = 0 to full scale, Zout=1KΩ |
3 | Current Monitor | 0-10V = 0 to full scale, Zout=1KΩ |
4 | Interlock Output | Connect 12V HV ON bulb to pin 15 to enable |
5 | +10V Reference | +10V at 1mA, maximum |
6 | Filament Monitor | 1V = 1A, Zout=1KΩ |
7 | Voltage Program Input | 0-10V = 0 to full scale, Zin=10MΩ |
8 | Local Voltage Program* | 0-10V, screwdriver adjust |
9 | Filament Limit Setpoint* | 1V = 1A, screwdriver adjust |
10 | Current Program Input | 0-10V = 0 to full scale, Zin=10MΩ |
11 | Local Current Program* | 10 turn pot, screwdriver adjust |
12 | Not used (+24V Out for Interlock) | (Optional Interlock configuration) |
13 | Not used (Interlock Coil) | (Optional Interlock configuration) |
14 | Filament Preheat Setpoint* | 1V = 1A, screwdriver adjust |
15 | Interlock Return | Interlock Ground |
*Denotes 10 turn potentiometer accessable through holes in cover
USB DIGITAL INTERFACE— 4 PIN USB “B” CONNECTOR
PIN | SIGNAL | SIGNAL PARAMETERS |
---|---|---|
1 | VBUS | +5Vdc |
2 | D- | Data - |
3 | D+ | Data + |
4 | GND | Ground |
HIGH VOLTAGE OUTPUT CONNECTOR
Spellman drywell type detachable connector.
50/65kV:
A one meter (39.4”) long polyethylene mating hig voltage cable with banana plug termination is provided.
80kV:
A one meter (39.4”) long polyethylene mating high voltage cable with corona ball termination is provided.
HV Cable Options:
5302:
(50kV unit only)
A one meter (39.4”) long Mammoflex mating high voltage
cable is provided, SHV p/n 201946-007
2001:
(50kV with XCC option and 65kV units only)
A one meter (39.4”) long Mammoflex mating high voltage cable
is provided, compatible with the XCC Option SHV p/n 201946-002
HIGH VOLTAGE POWER INPUT CONNECTOR 2 PIN PHOENIX CONTACT
PIN | SIGNAL | PARAMETER |
---|---|---|
1 | +24V High Voltage Power Input | +24V @ 7.75A max |
2 | +24V High Voltage Power Return | Power Ground |
ETHERNET DIGITAL INTERFACE— 8 PIN RJ45 CONNECTOR
PIN | SIGNAL | SIGNAL PARAMETERS |
---|---|---|
1 | TX+ | Transmit Data + |
2 | TX- | Transmit Data - |
3 | RX+ | Receive Data + |
4 | NC | No Connection |
5 | NC | No Connection |
6 | RX- | Receive Data - |
7 | NC | No Connection |
8 | NC | No Connection |
RS-232 DIGITAL INTERFACE— 9 PIN FEMALE D CONNECTOR
PIN | SIGNAL | SIGNAL PARAMETERS |
---|---|---|
1 | NC | No Connection |
2 | TX out | Transmit Data |
3 | RX in | Receive Data |
4 | NC | No Connection |
5 | SGND | Ground |
6 | NC | No Connection |
7 | NC | No Connection |
8 | Voltage Monitor 2 | 0-10V = 0 to full scale, Zout = 1KΩ |
9 | Power Supply OK | +15V = OK, 0V = Fault, Sink/Source 3mA max |
How to Order: |
---|
Sample model number: |
表和图
Digital Interface
The uXHP features a standard USB, RS-232 and Ethernet digital interface. Utilizing these standard digital interfaces can dramatically simplify power supply interfacing requirements saving the user both time and money, while enhancing functionality and overall capabiity. Spellman provides a GUI with the uXHP that allows the customer to both customize operational features of the uXHP while also providing basic power supply operational features. Details of the uXHP’s digital interface capability are described in detail in the uXHP manual.
Closeup showing digital interface connectors
Varex VF-80J X-Ray Tube
The uXHP was specifically designed in cooperation with Varex to power their new VF-80J X-Ray tube. This joint effort resulted in the creation of both the VF-80J X-Ray tube and the uXHP X-Ray generator. The Varex VF-80J tube is uniquely rated to run at up to 80kV @ 100 watts, 4mA maximum. The uXHP was explicitly designed to operate at up to 80kV and provide 100 watts down to 25kV, so customers could make full use of the capabilities of the VF-80J X-Ray tube. Varex and Spellman, working together to push the boundaries of X-Ray technology.
Frequently Asked Questions
Application Notes AN-12 – The Benefit of Using a Current Source to Power X-Ray Tube Filament Circuits
Application Notes AN-01 – Fundamentals of X-Ray Generator – X-Ray Tube Optimization